There are four primary variables used in this process. These are temperature, humidity, the time and gas concentration. In this process, a product which needs to be sterilized kept in a vacuum container. Then the product is exposed to ethylene oxide gas. Ethylene oxide gas is an alkylating agent that can disrupt the DNA of bacteria, micro-organisms thus stop them to reproduce in the product.
The ethylene oxide sterilizer is essentially a container that provides temperature, humidity, pressure, and time. It provides a perfect independent room for the ethylene oxide gas ,where EO gas kill all the bacterial deadly without corrosive to metal, rubber ,plastic and glass materials .
Control system | PLC unit and 7 “ touch screen |
Sterilization chamber | SS304 stainless steel |
Vacuum system | High-performance, low-noise, oil-free vacuum pump |
Built-in printer | Thermal printer |
Humidity system | Use saturated steam to humidify the sterilization chamber, and the air humidity is more uniform |
Timing system | EO gas diffusion time and residual EO aeration timing settable |
Pre-set | Standard Sterilization Mode and Residual Washing |
Model | SQ-H | SQ-E3 | SQ-E6 | SQ-E10 | SQ-E20 |
Capacity | ≤ 1 cbm | 3 cbm | 6 cbm | 10 cbm | 20 cbm |
Sterilizing temperature | 36-55 C adjustable | ||||
Humidity | 40-80% adjustable | ||||
Working pressure | -50 kpa | ||||
Sterilizing Time | 8-14 Hours , depends on materials | ||||
Sterilization Method | Low temperature , gas sterilization . | ||||
Display | LCD screen , touch screen | ||||
Safety Features | Temperature and Pressure protection | ||||
Accessory | Pallet and trolley | ||||
Voltage | 380V, 50Hz | ||||
Data Review | Thermal mini-printer and USB port (download) | ||||
EO gas | 80%EO +20%CO2 blend EO gas | ||||
Gas Consumption | 1kg/ cbm |
ethylene oxide sterilizer is ethylene oxide gas for the sterilization of media used for sterilization of damp-heat sensitive items, such as surgical precision-medical equipment, optical instruments, electronic equipment, plastic products, and various medical supplies, etc, for plastic, metal, and rubber materials will not produce any corrosion.
No. | Model | Volume | Chamber size(L*W*H) | Carton size(L*W*H) |
1 |
SQ-E3 |
3000L |
220*100*143cm |
245*170*205cm |
2 |
SQ-E4 |
4000L |
290*100*143cm |
315*170*205cm |
3 |
SQ-E5 |
5000L |
360*100*143cm |
385*170*205cm |
4 |
SQ-E6 |
6000L |
335*120*150cm |
355*175*205cm |
Medical sterilize disposables , surgical masks ,gown , swab ,syringes , infusion sets , Human implant consumables which can’t sterilized by steam autoclave and other sterilization method .